Journal article
Stent shortening and elongation: pitfalls with the Wiktor coronary endoprosthesis.
Published in:
- Catheterization and cardiovascular diagnosis. - 1994
English
This report demonstrates two specific complications that occurred during implantation of a Wiktor (Medtronic, Inc., Minneapolis, MN) coronary stent: stent shortening and elongation were observed and attributed to the unique coil structure of the device. Major cardiac events occurred during follow-up as a consequence of this stent deformation. Therefore, semiurgent bypass surgery should be considered to treat this complication.
-
Language
-
-
Open access status
-
closed
-
Identifiers
-
-
Persistent URL
-
https://sonar.rero.ch/global/documents/124165
Statistics
Document views: 7
File downloads: